The influence of rotational error and axial shift of toric intraocular lenses on residual astigmatism

Gargallo, Diana (Universidad de Zaragoza) ; Remón, Laura (Universidad de Zaragoza) ; Ares, Jorge (Universidad de Zaragoza) ; Castro-Alonso, Francisco J. (Universidad de Zaragoza)
The influence of rotational error and axial shift of toric intraocular lenses on residual astigmatism
Resumen: Purpose
Accurate alignment of Toric Intraocular Lens (T-IOLs) in cataract surgery is crucial for good visual outcomes. The purpose of this study was to evaluate the influence of rotation, axial shift and their combined effects on the refractive error and image quality of a wide range of T-IOL powers (from +1.50 D to +6.00 D cylinder) and two pupil diameters (3.34 and 4.44 mm).
Methods
Numerical ray tracing was utilized to quantify the residual error. Simulated retinal images and Visual Strehl (VS) ratios were calculated to evaluate image quality.
Results
First, T-IOL rotation showed better agreement with Holladay’s formula than 3.33% rule. Second, axial displacement resulted in acceptable residual cylinder (<0.50 D) across all examined cylinder powers. Third, concerning the combined effects, the influence of axial shift on residual cylinder becomes negligible when rotation errors exceed 2.5°. Fourth, a pupil-dependent nonlinear relationship was noted for image quality caused by both types of misalignment factors.
Conclusions
The 3.33% rule was confirmed as a reasonable approximation for the residual astigmatism caused by rotation of T-IOLs. The influence of axial shift on residual astigmatism becomes insignificant when there is also rotation. Image quality studies confirm that 30° of rotation are enough invalidate the compensation benefits of a T-IOLs in comparison with a Spherical Intraocular lens.

Idioma: Inglés
DOI: 10.1371/journal.pone.0311566
Año: 2024
Publicado en: PLoS ONE 19, 12 (2024), e0311566 [18 pp.]
ISSN: 1932-6203

Financiación: info:eu-repo/grantAgreement/ES/DGA-FEDER/E44-23R
Financiación: info:eu-repo/grantAgreement/ES/MICIU/PID2020-114311RA-I00
Tipo y forma: Artículo (Versión definitiva)
Área (Departamento): Área Oftalmología (Dpto. Cirugía)
Área (Departamento): Área Óptica (Dpto. Física Aplicada)


Creative Commons Debe reconocer adecuadamente la autoría, proporcionar un enlace a la licencia e indicar si se han realizado cambios. Puede hacerlo de cualquier manera razonable, pero no de una manera que sugiera que tiene el apoyo del licenciador o lo recibe por el uso que hace.


Exportado de SIDERAL (2025-01-03-13:21:43)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
Artículos > Artículos por área > Oftalmología
Artículos > Artículos por área > Optica



 Registro creado el 2025-01-03, última modificación el 2025-01-03


Versión publicada:
 PDF
Valore este documento:

Rate this document:
1
2
3
 
(Sin ninguna reseña)