Comparison of anterior segment measurements obtained by three different devices in healthy eyes
Resumen: Purpose. To assess the normal values and the repeatability of the Galilei Dual Scheimpflug Analyzer (GDSA), the biometer IOL Master, and the autokerato/refractometerWAM5500 in anterior segment examinations. Methods. Eighty-eight eyes from88healthy volunteers were prospectively and consecutively recruited. The repeatability was assessed, calculating the intraclass correlation coefficient (ICC). Results.The correlations among the repeated measurements showed nearly perfect reliability (ICC > 0.81) for all of the parameters, except corneal astigmatism Galilei (0.79) and WAM (0.68).There were statistically significant differences ( P < 0.001) between the values of the flat simulated keratometry (SimK) and the steep SimK measured by GDSA and the other methods; however, there were no statistically significant differences for the values obtained with the IOL Master andWAM 5500 ( P = 0.302 and P = 0.172, resp.) or between the values of the ACD( P < 0.001) andWTW( P = 0.007)measured by the IOLMaster and GDSA. Conclusions.The anterior segment measurements from the IOL Master and WAM 5500 were highly repeatable, comparable, and well correlated. In healthy young persons, the evaluated parameters had very good repeatability, although significant differences were found between the GDSA and IOL Master and between the GDSA and WAM 5500.
Idioma: Inglés
DOI: 10.1155/2014/498080
Año: 2014
Publicado en: BioMed Research International 2014 (2014), 498080 [8 pp]
ISSN: 2314-6133

Factor impacto JCR: 1.579 (2014)
Categ. JCR: MEDICINE, RESEARCH & EXPERIMENTAL rank: 85 / 123 = 0.691 (2014) - Q3 - T3
Categ. JCR: BIOTECHNOLOGY & APPLIED MICROBIOLOGY rank: 107 / 163 = 0.656 (2014) - Q3 - T2

Factor impacto SCIMAGO:

Financiación: info:eu-repo/grantAgreement/ES/MINECO/FIS2011-22496
Financiación: info:eu-repo/grantAgreement/ES/UZ/UZ2011-BIO-02
Tipo y forma: Artículo (Versión definitiva)
Área (Departamento): Oftalmología (Departamento de Cirugía, Ginecología y Obstetricia)

Creative Commons Debe reconocer adecuadamente la autoría, proporcionar un enlace a la licencia e indicar si se han realizado cambios. Puede hacerlo de cualquier manera razonable, pero no de una manera que sugiera que tiene el apoyo del licenciador o lo recibe por el uso que hace.


Exportado de SIDERAL (2017-05-18-10:42:26)

Este artículo se encuentra en las siguientes colecciones:
Artículos > Artículos por área > Oftalmología



 Registro creado el 2016-12-16, última modificación el 2017-05-18


Versión publicada:
 PDF
Valore este documento:

Rate this document:
1
2
3
 
(Sin ninguna reseña)