Effect of decentration, tilt and rotation on the optical quality of various toric intraocular lens designs: a numerical and experimental study
Resumen: Toric intraocular lenses (T-IOLs) may lose their optical quality if they are not correctly positioned inside the capsular bag once implanted. In this work, T-IOLs with cylinder powers of +1.50, +4.50 and +7.50 D and differing degrees of spherical aberration have been designed, manufactured and tested in vitro using a commercial optical bench that complies with the requirements of standard ISO 11979-2. Moreover, the effect of tilt and rotation on optical quality was assessed by means of numerical ray tracing on an astigmatic eye model, while the effect of decentration was evaluated numerically and experimentally.
Idioma: Inglés
DOI: 10.1364/BOE.447045
Año: 2022
Publicado en: Biomedical Optics Express 13, 4 (2022), 1948-1967
ISSN: 2156-7085

Financiación: info:eu-repo/grantAgreement/ES/DGA-FEDER/E44-20R
Financiación: info:eu-repo/grantAgreement/ES/DGA-FSE/T24-20R
Financiación: info:eu-repo/grantAgreement/ES/MCIU/DPI2017-84047-R
Financiación: info:eu-repo/grantAgreement/ES/MICIU/PID2020-114311RA-I00
Financiación: info:eu-repo/grantAgreement/ES/MINECO/DI-16-08888
Tipo y forma: Artículo (Versión definitiva)
Área (Departamento): Área Óptica (Dpto. Física Aplicada)

Derechos Reservados Derechos reservados por el editor de la revista


Exportado de SIDERAL (2022-05-11-12:48:02)

Este artículo se encuentra en las siguientes colecciones:
Artículos



 Registro creado el 2022-05-11, última modificación el 2022-05-11


Versión publicada:
 PDF
Valore este documento:

Rate this document:
1
2
3
 
(Sin ninguna reseña)