Reproducibility of retinal and choroidal measurements using swept-source optical coherence tomography in patients with Parkinson's disease
Resumen: PURPOSE: To assess the reproducibility of retinal and choroidal measurements in the macular and peripapillary areas using swept-source optical coherence tomography in patients with Parkinson''s disease. METHODS: A total of 63 eyes of 63 patients with idiopathic Parkinson''s disease were evaluated using a three-dimensional protocol of swept-source optical coherence tomography. The following layers were analyzed: full retinal thickness, retinal nerve fiber layer, ganglion cell layer, and choroid. The coefficient of variation was calculated for every measurement. RESULTS: In the macular area, the mean coefficients of variation of retinal thickness, ganglion cell layer + thickness, and choroidal thickness were 0.40%, 0.84%, and 2.09%, respectively. Regarding the peripapillary area, the mean coefficient of variation of the retinal nerve fiber layer thickness was 2.78. The inferior quadrant showed the highest reproducibility (coefficient of variation= 1.62%), whereas the superonasal sector showed the lowest reproducibility (coefficient of variation= 8.76%). CONCLUSIONS: Swept-source optical coherence tomography provides highly reproducible measurements of retinal and choroidal thickness in both the macular and peripapillary areas. The reproducibility is higher in measurements of retinal thickness versus choroidal thickness.
Idioma: Inglés
DOI: 10.5935/0004-2749.20200008
Año: 2020
Publicado en: ARQUIVOS BRASILEIROS DE OFTALMOLOGIA 83, 1 (2020), 19-27
ISSN: 0004-2749

Factor impacto JCR: 0.872 (2020)
Categ. JCR: OPHTHALMOLOGY rank: 60 / 62 = 0.968 (2020) - Q4 - T3
Factor impacto SCIMAGO: 0.602 - Ophthalmology (Q2) - Medicine (miscellaneous) (Q2)

Tipo y forma: Artículo (Versión definitiva)
Área (Departamento): Área Oftalmología (Dpto. Cirugía)
Área (Departamento): Área Óptica (Dpto. Física Aplicada)


Creative Commons Debe reconocer adecuadamente la autoría, proporcionar un enlace a la licencia e indicar si se han realizado cambios. Puede hacerlo de cualquier manera razonable, pero no de una manera que sugiera que tiene el apoyo del licenciador o lo recibe por el uso que hace.


Exportado de SIDERAL (2025-04-10-14:04:29)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
Artículos > Artículos por área > Oftalmología
Artículos > Artículos por área > Optica



 Registro creado el 2025-04-10, última modificación el 2025-04-10


Versión publicada:
 PDF
Valore este documento:

Rate this document:
1
2
3
 
(Sin ninguna reseña)