Versatile lock and key assembly for optical measurements with microfluidic platforms and cartridges
Resumen: A novel and versatile optical reader for microfluidic platforms is presented. The reader includes a modular insertion port based on the lock and key concept for reproducible alignment with a miniaturized optical detection system comprising an interchangeable light emitting diode (LED) and a photodiode. The modular nature of the insertion port allows the use of microfluidic platforms in variable shapes and fluidic configurations. Three different analytical methodologies based on absorbance or fluorescence measurements were used to demonstrate the flexibility and reproducibility of the proposed experimental setup.
Idioma: Inglés
DOI: 10.1021/ac504255t
Año: 2015
Publicado en: ANALYTICAL CHEMISTRY 87, 3 (2015), 1503-1508
ISSN: 0003-2700

Factor impacto JCR: 5.886 (2015)
Categ. JCR: CHEMISTRY, ANALYTICAL rank: 4 / 75 = 0.053 (2015) - Q1 - T1
Factor impacto SCIMAGO: 2.241 - Analytical Chemistry (Q1)

Tipo y forma: Article (Published version)

Rights Reserved All rights reserved by journal editor