Wide-field direct ocular straylight meter

Ávila, Francisco (Universidad de Zaragoza) ; Collados, María Victoria (Universidad de Zaragoza) ; Ares, Jorge (Universidad de Zaragoza) ; Remón, Laura (Universidad de Zaragoza)
Wide-field direct ocular straylight meter
Resumen: The impact of the intraocular straylight (IOS) on the visual performance and retinal imaging is still a challenging topic. Direct optical methods to measure IOS avoid psychophysical approaches and interaction with the patient. In this work, we developed an optical instrument providing direct imaging measurement of IOS based on the double-pass technology. The system was tested in an artificial eye IOS model constructed with holographic diffusers and validated with theoretical simulations.
Idioma: Inglés
DOI: 10.1364/OE.387940
Año: 2020
Publicado en: OPTICS EXPRESS 28, 8 (2020), 11237-11242
ISSN: 1094-4087

Factor impacto JCR: 3.894 (2020)
Categ. JCR: OPTICS rank: 20 / 99 = 0.202 (2020) - Q1 - T1
Factor impacto SCIMAGO: 1.393 - Atomic and Molecular Physics, and Optics (Q1)

Financiación: info:eu-repo/grantAgreement/ES/DGA-FEDER/E24-17R
Financiación: info:eu-repo/grantAgreement/ES/DGA/R44-17R
Financiación: info:eu-repo/grantAgreement/ES/MINECO/DPI2017-84047-R
Tipo y forma: Article (Published version)
Área (Departamento): Área Óptica (Dpto. Física Aplicada)
Exportado de SIDERAL (2023-12-19-14:44:40)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
articulos > articulos-por-area > optica



 Notice créée le 2020-06-16, modifiée le 2023-12-19


Versión publicada:
 PDF
Évaluer ce document:

Rate this document:
1
2
3
 
(Pas encore évalué)