Novel electronic device to quantify the cyclic fatigue resistance of endodontic reciprocating files after using and sterilization
Resumen: Background: The aim of this study was to analyze the effects of the time of use (TU) and sterilization cycles (SC) of endodontic reciprocating files on cyclic fatigue resistance. Methods: One-hundred-and-twenty (120) Procodile NiTi endodontic reciprocating instruments were selected at random and distributed into the following study groups: A: 0 sterilization cycles/0s time of use (n = 10); B: 0/60 (n = 10); C: 0/120 (n = 10); D: 1/0 (n = 10); E: 1/60 (n = 10); F: 1/120 (n = 10); G: 5/0 (n = 10); H: 5/60 (n = 10); I: 5/120 (n = 10); J: 10/0 (n = 10); K: 10/60 (n = 10); and L: 10/120 (n = 10). A dynamic cyclic fatigue device was designed using computer-aided design/computer-aided engineering (CAD/CAE) technology and created with a 3D printer to simulate the pecking motion performed by the clinician. Failure of the endodontic rotary instrument was detected by a light-emitting diode-light-dependent resistor (LED-LDR) system controlled by an Arduino driver complex and management software. The results were analyzed using the ANOVA test. Results: All pairwise comparisons presented statistically significant differences between the time to failure, number of cycles to failure and number of cycles of in-and-out movement for the time of use study groups (p < 0.001), but not in the number of sterilization cycles (p > 0.05). Conclusions: The time of use of NiTi endodontic reciprocating files negatively affects dynamic cyclic fatigue resistance. Dynamic cyclic resistance is not affected by the number of sterilization cycles.
Idioma: Inglés
DOI: 10.3390/app10144962
Año: 2020
Publicado en: APPLIED SCIENCES-BASEL 10, 14 (2020), 4962 [14 pp.]
ISSN: 2076-3417

Factor impacto JCR: 2.679 (2020)
Categ. JCR: PHYSICS, APPLIED rank: 73 / 160 = 0.456 (2020) - Q2 - T2
Categ. JCR: ENGINEERING, MULTIDISCIPLINARY rank: 38 / 91 = 0.418 (2020) - Q2 - T2
Categ. JCR: CHEMISTRY, MULTIDISCIPLINARY rank: 101 / 178 = 0.567 (2020) - Q3 - T2
Categ. JCR: MATERIALS SCIENCE, MULTIDISCIPLINARY rank: 201 / 333 = 0.604 (2020) - Q3 - T2

Factor impacto SCIMAGO: 0.435 - Computer Science Applications (Q2) - Engineering (miscellaneous) (Q2) - Process Chemistry and Technology (Q2) - Instrumentation (Q2) - Materials Science (miscellaneous) (Q2) - Fluid Flow and Transfer Processes (Q2)

Tipo y forma: Artículo (Versión definitiva)
Área (Departamento): Área Estomatología (Dpto. Cirugía)

Creative Commons Debe reconocer adecuadamente la autoría, proporcionar un enlace a la licencia e indicar si se han realizado cambios. Puede hacerlo de cualquier manera razonable, pero no de una manera que sugiera que tiene el apoyo del licenciador o lo recibe por el uso que hace.


Exportado de SIDERAL (2021-09-02-10:10:04)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
Artículos



 Registro creado el 2020-09-10, última modificación el 2021-09-02


Versión publicada:
 PDF
Valore este documento:

Rate this document:
1
2
3
 
(Sin ninguna reseña)