Biomechanically-driven simulations of the MyoRing treatment in subjects with high myopia
Financiación H2020 / H2020 Funds
Resumen: Purpose: Corneal biomechanics is a determinant factor for the outcome of continuous intracorneal segments (MyoRing, Dioptex GmbH). However, implant selection remains driven by optical parameters such as the average central curvature of the cornea (Kmean) and the Spherical Equivalent (SE), while failing to account for the tissue biomechanics or the intraocular pressure (IOP). We hypothesize that biomechanical models and computer simulation can improve the refractive outcomes of the MyoRing treatment.

Methods: Four thousand computer models representing the population of patients who are candidates for a MyoRing treatment have been created using the finite element method. These numerical models accounted for physiologic variability of the anatomical features (anterior and posterior corneal radius of curvature, corneal thickness, axial length and pupil size) and mechanical properties (corneal biomechanics and IOP). Two MyoRing implants were evaluated on these virtual patients (Myoring size: 280 µm; optical zone: 5 and 6 mm; depth: 60%-75%; laser ablated pocket 5 µm). Refractive outcomes obtained with the biomechanical simulation were compared to previous clinical data (Daxer 2017; Rattan 2018).

Results: Population-based simulations were able to reproduce the refractive correction observed clinically; the average post-surgical Kmean was within ± 0.5 D of the data reported by Rattan (Fig.1 a), and the distribution of post-surgical SE closely matched the clinical data from Daxer (Fig.1 b). Moreover, our simulation approach allowed us to determine the clinical parameters having the most important contributions to the refractive outcome. For example, we found that the correction in Kmean is strongly related to the pre-surgical thickness, mechanics, and IOP, while the correction in SE is highly affected by the depth of MyoRing insertion.

Conclusions: Biomechanical modelling is able to predict the refractive outcomes of MyoRing implantations. This approach provides a deeper understanding on the mechanisms underlying continuous intracorneal segments, which can be used to improve implant design and further personalize the treatment procedure.

Idioma: Inglés
Año: 2020
Publicado en: Investigative ophthalmology & visual science 61, 7 (2020), 4722 [3 pp]
ISSN: 0146-0404

Originalmente disponible en: Texto completo de la revista

Factor impacto JCR: 4.799 (2020)
Categ. JCR: OPHTHALMOLOGY rank: 8 / 62 = 0.129 (2020) - Q1 - T1
Factor impacto SCIMAGO: 1.935 - Cellular and Molecular Neuroscience (Q1) - Sensory Systems (Q1) - Ophthalmology (Q1)

Financiación: info:eu-repo/grantAgreement/EC/H2020/786692/EU/Multiscale Integrative Approach for Corneal Biomechanics to Assess Corneal Crosslinking/MIMetiCO
Tipo y forma: Article (Published version)
Área (Departamento): Área Mec.Med.Cont. y Teor.Est. (Dpto. Ingeniería Mecánica)
Exportado de SIDERAL (2022-04-25-10:39:25)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
articulos



 Notice créée le 2020-10-28, modifiée le 2022-04-25


Versión publicada:
 PDF
Évaluer ce document:

Rate this document:
1
2
3
 
(Pas encore évalué)