A first test of CUPID prototypal light detectors with NTD-Ge sensors in a pulse-tube cryostat

Alfonso, K. ; Armatol, A. ; Augier, C. ; Avignone, F.T. ; Azzolini, O. ; Balata, M. ; Barabash, A.S. ; Bari, G. ; Barresi, A. ; Baudin, D. ; Bellini, F. ; Benato, G. ; Berest, V. ; Beretta, M. ; Bettelli, M. ; Biassoni, M. ; Billard, J. ; Boldrini, V. ; Branca, A. ; Brofferio, C. ; Bucci, C. ; Camilleri, J. ; Campani, A. ; Capelli, C. ; Capelli, S. ; Cappelli, L. ; Cardani, L. ; Carniti, P. ; Casali, N. ; Celi, E. ; Chang, C. ; Chiesa, D. ; Clemenza, M. ; Colantoni, I. ; Copello, S. ; Craft, E. ; Cremonesi, O. ; Creswick, R.J. ; Cruciani, A. ; D'Addabbo, A. ; D'Imperio, G. ; Dabagov, S. ; Dafinei, I. ; Danevich, F.A. ; De Jesus, M. ; de Marcillac, P. ; Dell'Oro, S. ; Di Domizio, S. ; Di Lorenzo, S. ; Dixon, T. ; Dompé, V. ; Drobizhev, A. ; Dumoulin, L. ; Fantini, G. ; Faverzani, M. ; Ferri, E. ; Ferri, F. ; Ferroni, F. ; Figueroa-Feliciano, E. ; Foggetta, L. ; Formaggio, J. ; Franceschi, A. ; Fu, C. ; Fu, S. ; Fujikawa, B.K. ; Gallas, A. ; Gascon, J. ; Ghislandi, S. ; Giachero, A. ; Gianvecchio, A. ; Girola, M. ; Gironi, L. ; Giuliani, A. ; Gorla, P. ; Gotti, C. ; Grant, C. ; Gras, P. ; Guillaumon, P.V. ; Gutierrez, T.D. ; Han, K. ; Hansen, E.V. ; Heeger, K.M. ; Helis, D.L. ; Huang, H.Z. ; Imbert, L. ; Johnston, J. ; Juillard, A. ; Karapetrov, G. ; Keppel, G. ; Khalife, H. ; Kobychev, V.V. ; Kolomensky, Yu.G. ; Konovalov, S.I. ; Kowalski, R. ; Langford, T. ; Lefevre, M. ; Liu, R. ; Liu, Y. ; Loaiza, P. ; Ma, L. ; Madhukuttan, M. ; Mancarella, F. ; Marini, L. ; Marnieros, S. ; Martinez, M. (Universidad de Zaragoza) ; Maruyama, R.H. ; Mas, Ph. ; Mayer, D. ; Mazzitelli, G. ; Mei, Y. ; Milana, S. ; Morganti, S. ; Napolitano, T. ; Nastasi, M. ; Nikkel, J. ; Nisi, S. ; Nones, C. ; Norman, E.B. ; Novosad, V. ; Nutini, I. ; O'Donnell, T. ; Olivieri, E. ; Olmi, M. ; Ouellet, J.L. ; Pagan, S. ; Pagliarone, C. ; Pagnanini, L. ; Pattavina, L. ; Pavan, M. ; Peng, H. ; Pessina, G. ; Pettinacci, V. ; Pira, C. ; Pirro, S. ; Poda, D.V. ; Polischuk, O.G. ; Ponce, I. ; Pozzi, S. ; Previtali, E. ; Puiu, A. ; Quitadamo, S. ; Ressa, A. ; Rizzoli, R. ; Rosenfeld, C. ; Rosier, P. ; Scarpaci, J.A. ; Schmidt, B. ; Sharma, V. ; Shlegel, V.N. ; Singh, V. ; Sisti, M. ; Slocum, P. ; Speller, D. ; Surukuchi, P.T. ; Taffarello, L. ; Tomei, C. ; Torres, J.A. ; Tretyak, V.I. ; Tsymbaliuk, A. ; Velazquez, M. ; Vetter, K.J. ; Wagaarachchi, S.L. ; Wang, G. ; Wang, L. ; Wang, R. ; Welliver, B. ; Wilson, J. ; Wilson, K. ; Winslow, L.A. ; Xue, M. ; Yan, L. ; Yang, J. ; Yefremenko, V. ; Umatov, V.I. ; Zarytskyy, M.M. ; Zhang, J. ; Zolotarova, A. ; Zucchelli, S.
A first test of CUPID prototypal light detectors with NTD-Ge sensors in a pulse-tube cryostat
Financiación H2020 / H2020 Funds
Resumen: CUPID is a next-generation bolometric experiment aiming at searching for neutrinoless double-beta decay with ∼250 kg of isotopic mass of 100Mo. It will operate at ∼10 mK in a cryostat currently hosting a similar-scale bolometric array for the CUORE experiment at the Gran Sasso National Laboratory (Italy). CUPID will be based on large-volume scintillating bolometers consisting of 100Mo-enriched Li2MoO4 crystals, facing thin Ge-wafer-based bolometric light detectors. In the CUPID design, the detector structure is novel and needs to be validated. In particular, the CUORE cryostat presents a high level of mechanical vibrations due to the use of pulse tubes and the effect of vibrations on the detector performance must be investigated. In this paper we report the first test of the CUPID-design bolometric light detectors with NTD-Ge sensors in a dilution refrigerator equipped with a pulse tube in an above-ground lab. Light detectors are characterized in terms of sensitivity, energy resolution, pulse time constants, and noise power spectrum. Despite the challenging noisy environment due to pulse-tube-induced vibrations, we demonstrate that all the four tested light detectors comply with the CUPID goal in terms of intrinsic energy resolution of 100 eV RMS baseline noise. Indeed, we have measured 70–90 eV RMS for the four devices, which show an excellent reproducibility. We have also obtained high energy resolutions at the 356 keV line from a 133Ba source, as good as Ge semiconductor γ detectors in this energy range.
Idioma: Inglés
DOI: 10.1088/1748-0221/18/06/P06033
Año: 2023
Publicado en: Journal of Instrumentation 18, 6 (2023), P06033 [20 pp.]
ISSN: 1748-0221

Factor impacto JCR: 1.3 (2023)
Categ. JCR: INSTRUMENTS & INSTRUMENTATION rank: 57 / 76 = 0.75 (2023) - Q3 - T3
Factor impacto CITESCORE: 2.4 - Mathematical Physics (Q2) - Instrumentation (Q3)

Factor impacto SCIMAGO: 0.58 - Mathematical Physics (Q2) - Instrumentation (Q2)

Financiación: info:eu-repo/grantAgreement/EC/H2020/754496/EU/FELLowship for Innovation at INFN/FELLINI
Tipo y forma: Article (Published version)
Área (Departamento): Área Física Atóm.Molec.y Nucl. (Dpto. Física Teórica)
Exportado de SIDERAL (2024-11-22-12:09:39)


Visitas y descargas

Este artículo se encuentra en las siguientes colecciones:
articulos > articulos-por-area > fisica_atomica,_molecular_y_nuclear



 Notice créée le 2024-02-19, modifiée le 2024-11-25


Versión publicada:
 PDF
Évaluer ce document:

Rate this document:
1
2
3
 
(Pas encore évalué)